Removal and Replacement of Exhalation Flow Sensor Cable(outside) CBL2123P
1 Perform above steps for removal of Exhalation Valve and Exhalation
Flow Sensor.
2 Remove the Exhalation Flow Sensor Cable (outside) CBL2123P from
the Exhalation Manifold MNF2110A by pulling the collet of the cable
connector.
3 To reinstall the cable to the Exhalation Manifold MNF2110A, align
the key on the connector with the notch of the female connector on
the Exhalation Manifold MNF2110A.
Note: The lock mechanism of the connector can be easily released by
pulling the collet of the connector. Do NOT disconnect the connector by
pulling the wire.
Removal and Replacement of Top Cover CVR2101M
Figure 4-2 T
op Cover
Refer to figure 4-2 and follow these steps:
1 Rotate the e360 so the rear of the ventilator is facing you.
2 Locate the 6 screws securing the cover to the e360 and
remove.
SECTION 4
Remove 6
screws, 3
each side,
securing top
cover
SER360 A1106 4-3