EasyManua.ls Logo

Newport e360 - Setup

Newport e360
129 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1/4 inch Open-ended Wrench
5/16 inch Open-ended Wrench
Medium size flat head screw driver
Caution The accuracy of all test equipment (electronic or
pneumatic) used for verification or calibration procedures
should be certified annually by a testing laboratory before
use.
SETUP
Figure 5-1 e360 Ventilator Connected to Gas Sources
1 Visually inspect the e360 ventilator for cosmetic defects,
damage or missing items.
2 Before removing the Top Cover, unplug the e360 ventilator
from the AC power and gas sources.
3 Remove the 6 screws on the Back Panel that secures the
T
op
Cover to the e360 ventilator. Remove the Top Cover, see
Section 4, “Removal of Top Cover”.
4 Reconnect the AC power to the e360 ventilator.
5 Connect both air and oxygen supply gas sources to the e360
ventilator, verify supply gas sources are at 50 ±2 psig.
6 Prior to starting the calibration procedure, verify that all test
equipment is in current calibration status.
CALIBRATION PROCEDURES
O
xygen Supply
Air Supply
5-2 SER360 A1106

Table of Contents

Other manuals for Newport e360

Related product manuals