Figure 4-25 Exhalation Flow Sensor Board PCB2103A
Removal and Replacement of Battery BAT1800P
Figure 4-26 Battery removal BAT1800P
Refer to Figure 4-26 and perform these steps:
1 Perform steps for removal of Top Cover on Page 4-3.
2 Remove the battery wires from the positive and negative
terminals.
3 Remove the screw securing the battery bracket to the e360
ventilator and life off bracket.
4 Remove the battery by lifting up and out.
5 T
o reinstall, reverse steps 2 through 4.
SECTION 4
Remove screw
Positive terminal
(Red wire)
Negative terminal
(Black wire)
CBL2124P connected to
K5 on PCB2103P
CBL2122A connected to
K1 on PCB2103P
Polarization hole for K1
on PCB2103P
SER360 A1106 4-25