Removal and Replacement of AC/DC Power Supply PWR2100P
Figure 4-27 AC/DC Power Supply PWR2100P
Refer to figure 4-27 and follows these steps:
1 Perform above steps for removal of the battery
.
2 Removal cables CBL2115A from connector TB1 and
CBL2103A
from connector TB2 on PWR2100P
3 Remove 4 screws and washers securing PWR2100P to the
e360 ventilator.
4 Remove the PWR2100P.
5 To reinstall, reverse above steps 2 through 4.
COMPONENT REMOVAL & REPLACEMENT
Cable CBL2115
c
onnected to TB1
Remove 4 screws
a
nd washers
Cable CBL2103
c
onnected to TB2
4-26 SER360 A1106