Removal and Replacement of DC to DC Power Supply PCB2101A
Figure 4-28 Dc to DC power supply PCB2101P
Refer to figure 4-28 and follow steps:
1 Perform above steps for removal of battery BAT1880P.
2 Remove cables from the following locations;
• CBL2101A from J103 on PCB2101P
• CBL2103A from J104 on PCB2101P
• CBL2108A from J106 on PCB2101P
•
CBL21
1
1A
from J101 on PCB2101P
• CBL2119A from J102 on PCB2101P
3 Remove 4 screws and washers securing PCB2101P to the
e360 ventilator.
4 T
o reinstall, reverse above steps 2 and 3.
SECTION 4
CBL2103A at J104 CBL2111A at J101 CBL2101A at J103 CN1 on PCB2106P
Remove 4 screws
and washers
CBL2108A at J106
CBL2119A at J102
SER360 A1106 4-27