Removal and Replacement of Exhalation Manifold MNF2110A
Refer to figures 4-23 and 4-24 and follow these steps:
1 Turn the e360 ventilator so the bottom of the unit is facing
upward, remove 3 screws securing the MNF2101A to the base
of the e360 and the 1 screw and washer securing the bumper
BMP2100P to the bottom of the e360 ventilator.
2 Remove the exhalation valve drive line and the Machine 2
pressure line from the rear of the Exhalation Manifold
MNF2110A (see figure 4-22).
3 Remove the MNF2110A from the e360 ventilator.
4 To reinstall, reverse above steps 2 and 3.
Figure 4-23 Location of 3 screws and bumper
SECTION 4
Bumper BMP2100P Remove 3 screws
SER360 A1106 4-23