D1: Air Servo Valve Calibration
ID Meaning
4 Hit Manual Inflation Button to Start Air Servo Valve Calibration
5 Calibrating Air Servo Valve
6 Air Servo Valve Calibration Completed
7 Calibration Failed: Valve Leaks more than 50 cc
D2: Oxygen Servo Valve Calibration
ID Meaning
8 Hit Manual Inflation Button to Start Oxygen Servo Valve Calibration
9 Calibrating Oxygen Servo Valve
10
Oxygen Servo Valve Calibration Completed
7 Calibration Failed: Valve Leaks more than 50 cc
D3: Manual Flow Adjustment
ID Meaning
11 Hit Manual Inflation Button to Enable Manual Flow Adjustment
12 Adjust Flow Setting
13
Manual Flow Adjustment Disabled
D4: Exhalation Valve Calibration
ID Meaning
14 Hit Manual Inflation Button to Enable Exhalation Valve Calibration
35 Occlude the Patient Circuit and Hit Manual Inflation Button Again to Start Exhalation Valve
Calibration
15 Calibrating Exhalation Valve
20 Exhalation Valve Calibration Completed
16
Exhalation V
alve Calibration Failed: Proximal Pressure Less than 1cmH2O when Exhalation
Valve Closed
17 Exhalation Valve Calibration Failed: Proximal Pressure High than 1cmH2O when Exhalation
V
alve Opened
18 Exhalation Valve Calibration Failed: Proximal Pressure Low than Target Pressure
19
Exhalation V
alve Calibration Failed: Test Flow Less than 1L/min
D5: Analog Board Calibration
ID Meaning
21
Hit Manual Inflation Button to Start
Analog Board Calibration
25 Analog Board Calibration Enabled
26
Analog Board Calibration Disabled
SER360 A1106 5-15
SECTION 5
Table 5-6 Descriptions of Message ID