5
6-14 SER360 A1106
Inspiratory Time (sec) Ptrig Sensitivity (cmH
2
O)
Set t
I
Measured Tol. Set Actual Tol. Range
Insp. Time Range
0.25 0.2 - 0.3 1 + 0 - 2
0.5 0.4 - 0.6 3 +
2 - 4
Respiratory Rate (b/min) PEEP/CPAP (cmH
2
O)
Set ƒ
tot
Measured Tol. Set PEEP Measured Tol. Range
Resp. Rate Range Pressure
5+13 +1 Variance
Variance
10 +
1 Variance
20 +
120 +1 Variance
V
ariance
Pressure Control (cmH
2
O)
Inspiratory Tidal Volume (mL)
Set VTI Measured
Tol. Range
Volume
200 + 10%
Variance
500
+
10%
Variance
Pressure Support (cmH
2
O) Expiratory Tidal Volume (mL)
Set
V
T
E Tol.
Measured
Tol.
Range Volume Range
250 212-259 + 10%
Variance
500 423-517 +
10%
Variance
1000 880-1030 +
10%
V
ariance
OPERATIONAL VERIFICATION PROCEDURE
e360 OVP TEST RECORD SHEET (CONTINUED)
Set Ppeak Measured Tol.
Range
10
+
10%
Variance
20
+
10%
Variance
50
+
10%
Variance
Set
P
peak Measured
Tol.
Pressure Range
10
+
10%
V
ariance
20
+
10%
Variance