Device Alert Violation Priority Definitions
Violation Messages Level
Monitor ROM failure Monitor ROM
Failed
High Read only memory that stores the code of the
monitor processor has an incorrect checksum.
Monitor task continuity
failure
Mon Task Failed High Software tasks of the monitor processor have
operated out of sequence.
Power Failure Power Failure High DC power out of tolerance. Check +12 VDC,
–12 VDC and +5 VDC. (The e360 may have
been powered by internal battery until it was
depleted and a Device Alert resulted.)
TROUBLESHOOTING GUIDE
3-4 SER360 A1106