6.6 Performance Testing
6-10 ZM2376
Measurement frequency 10kHz, Measurement signal level 1 Vrms, Cable length 0m
Measurement
range
Standard
(nominal
value)
Calibration value of
standard
A
Measured value
B
Difference
100×(B-A)/A
Specification
B-A
1M
10pF
C F
C F
C %
C ± 0.86 %
D
D
D
D ± 0.0086
100k
100pF
C F
C F
C %
C ± 0.16 %
D
D
D
D ± 0.0016
10k
1000pF
C F
C F
C %
C ± 0.11 %
D
D
D
D ± 0.0011
1k
0.0F
C F
C F
C %
C ± 0.098 %
D
D
D
D ± 0.0010
100
0.F
C F
C F
C %
C ± 0.11 %
D
D
D
D ± 0.0011
10
10
Z
Z
Z %
Z ± 0.20 %
θ °
θ °
θ °
θ ± 0.12 °
Measurement frequency 100kHz, Measurement signal level 1 Vrms, Cable length 0m
Measurement
range
Standard
(nominal
value)
Calibration value of
standard
A
Measured value
B
Difference
100×(B-A)/A
Specification
B-A
100k
100k
Z
Z
Z %
Z ± 0.33 %
θ °
θ °
θ °
θ ± 0.19 °
10k
10k
Z
Z
Z %
Z ± 0.15 %
θ °
θ °
θ °
θ± 0.083 °
1k
1k
Z
Z
Z %
Z ± 0.14 %
θ °
θ °
θ °
θ± 0.081 °
100k
10pF
C F
C F
C %
C ± 0.41 %
D
D
D
D ± 0.0041
10k
100pF
C F
C F
C %
C ± 0.17 %
D
D
D
D ± 0.0017
1k
1000pF
C F
C F
C %
C ± 0.16 %
D
D
D
D ± 0.0016
100
0.0F
C F
C F
C %
C ± 0.11 %
D
D
D
D ± 0.0011
10
10
Z
Z
Z %
Z ± 0.23 %
θ °
θ °
θ °
θ± 0.13 °
For the 1k, 10k, and 100k ranges, either a standard resistor or a standard capacitor should be
used for the test.

Table of Contents