7.1 Specification
7-14 ZM2376
• Monitor display Voltage: Voltage value applied to the DUT
Voltage Monitor Accuracy
±(2%+2mVrms) from 10Hz to 50kHz
±(3.5%+2mVrms) over 50kHz to 100kHz
±(5%+5mVrms) over 100kHz to 1MHz
±(10%+10mVrms) over 1MHz to 5.5MHz
Current: Current value flowing in the DUT
Current monitor accuracy (supplementary value)
Voltage monitor accuracy + Measurement accuracy of
impedance Z
• Discharge protection 4J or less when voltage is below 250V, or 0.5J or less when below 1kV.
(All are supplementary values)
For voltage V[V], the energy stored in capacitance C[F] is (1/2) × C × V
2
[J].
Remote control interface
• USB USBTMC, USB 1.1 Full-speed
• RS-232 Data rate
4800 / 9600 / 19200 / 38400 / 57600 / 115200 / 230400bps
For the data rate exceeding 19200bps, communication may fail
depending on the characteristics of cable or controller.
Flow control
None, Software (X-ON/X-OFF), Hardware (RTS/CTS)
• GPIB Conforms to IEEE 488.1 and IEEE 488.2 Standards
• LAN (optional) 10BASE-T / 100BASE-TX, RJ-45 connector