7.1 Specification
7-13 ZM2376
Other measurement related functions
Zero correction OPEN correction and SHORT correction provided. Both can be turned on or
off.
LOAD correction Provided. It can be turned on or off.
Cable length correction 0m / 1m / 2m / 4m
Contact check Provided.
Based on detection of an abnormally low capacitance or abnormal
voltage/current.
Averaging 1 to 256 times
Deviation measurement Primary parameters: Deviation and deviation % from reference value can be
displayed.
Secondary parameters: Deviation and deviation % from reference value can be
displayed.
Comparator Primary parameters: Max. 14 bins
Original measured value / Deviation / Deviation % can
be sorted.
Secondary parameters: Upper limit and lower limit comparison
Original measured value / Deviation / Deviation % can
be sorted.
Beeper: Sounds according to comparison result
(Pass / Fail / Off)
Handler interface Signal isolation: All I/O signals are optically isolated
(withstand voltage ±42V)
Input signal: Trigger, Key lock, Settings/correction value memory
designation.
High speed recalling with only spot correction value is
possible.
Output signal: Comparison result BIN1 to BIN11, NC / BIN12, PHI /
BIN13, PLO / BIN14, OUT OF BINS, S-NG, ERR,
INDEX, EOM (when BIN10 - BIN14 are used, NC,
PHI, and PLO cannot be used).
Rated power voltage: External +5V to +24V, Internal +5V (non-isolated)
Multi-measurement Execute measurement and limit comparison under multiple conditions for
the total comparison.
Maximum number of steps: 32
Selectable measurement conditions: Measurement frequency, measurement
signal level, internal DC bias voltage, measurement parameters, etc.