The following codes can be in the type code column:
Type code Description Comments
0 Hardware reset Turning ON
1 Software reset Device restart from menu
2 Entry of measured value Programmed TREND logging entry
3 MultiCONT CRC altered
Settings of MultiCONT have changed. CRC calculated when entering
Main menu differs from the CRC calculated when quitting the Main
menu.
4 TOT1 clear Clearing TOT1 from a NIVELCO instrument used for flow-metering.
5 TOT2 clear Clearing TOT2 from a NIVELCO instrument used for flow-metering.
6 MultiCONT clock set
7 Transmitter ( DEVICE ) default load Loading default settings of the transmitter from menu
8 MultiCONT default load
9 Logger delete from menu
10 MultiCONT error-list clear
11 Transmitter program CRC altered
MultiCONT checks CRC when entering and quitting remote
programming mode. Logging entries are created when the two CRC
differs.
12 New list saved after ( DEVICE ) detection
253 Instrument respond error
Transmitter does not respond to the entered commands, it will be stored
also in the error-list
254 Instrument inactive The instrument selected for logging is in inactive state