1-4
CHAPTER 1 INTRODUCTION
Capacity (g) 210 410 100/410 610 2100 4100 1000/4100 4100 6100 8100
Readability (g) 0.001 0.001/0.01 0.01 0.01/0.1 0.1
Repeatability (Std. dev.) (g) 0.0005 0.0005/0.005 0.005 0.01/0.05 0.05
Linearity (g) (+) 0.002 0.002/0.005 0.02 0.02/0.05 0.1
Tare range Full capacity by subtraction
Stabilization time <3 seconds
Sensitivity drift PPM/°C 4 3 4 3 4 3 4 3
(10°C - 30°C)
Operating temperature range: ..................
w/internal calibration 10° to 40°C/50° to 104°F ..................
w/o internal calibration 10° to 30°C/50° to 86°F ..................
Power requirements External Adapter, 100-120 V ac, 220 V ac, 50/60 Hz
Plug configuration for US, Euro, UK & Australia
TABLE 1-3. PRECISION TOP LOADER SPECIFICATIONS
Capacity (g) 62 110 210 100/210
Readability (mg) 0.1 0.1/1
Repeatability (Std. dev.) (mg) 0.1 0.1/0.5
Linearity (mg) (+) 0.2 0.2/0.5
Tare range Full capacity by subtraction
Safe overload capacity 150% of capacity
Stabilization time <4 seconds
Sensitivity drift PPM/°C 3 ..................
(10°C - 30°C)
Operating temperature range: ..................
w/internal calibration 10° to 40°C/ 50° to 104°F ..................
w/o internal calibration 10° to 30°C/50° to 86°F
Power requirements External Adapter, 100-120 V ac, 220 V ac, 50/60 Hz
Plug configuration for US, Euro, UK & Australia
1.5 SPECIFICATIONS
Complete specificatons for the Ohaus Explorer balances are listed in Tables 1-2, 1-3 and 1-4. When a balance has been
serviced, it must meet the specifications listed in the table. Before servicing the balance, determine what specifications
are not met.
Capacity (g) 12,000 22,000 32,0000
Readability (g) 0.1
Repeatability (Std. dev.) (g) 0.1
Linearity (g) (+) 0.4
Tare range Full capacity by subtraction
Safe overload capacity 150% of capacity
Stabilization time <4 seconds
Sensitivity drift PPM/°C 3 ..................
(10°C - 30°C)
Operating temperature range: ..................
w/internal calibration 10° to 40°C/ 50° to 104°F ..................
w/o internal calibration 10° to 30°C/50° to 86°F
Power requirements External Adapter, 100-120 V ac, 220 V ac, 50/60 Hz
Plug configuration for US, Euro, UK, Japan & Australia
TABLE 1-4. HIGH CAPACITY TOP LOADER SPECIFICATIONS
TABLE 1-2. ANALYTICAL SPECIFICATIONS