2-6
CHAPTER 2 DIAGNOSIS
TABLE 2-4. EXPLORER PRO ERROR CODES (Cont.)
Displays Err 9.4
Factory internal calibration data failed checksum. Perform MFG and MFGWT. If the problem persists reload the TC data.
If the problem still persists the main PCB must be reconfigured at Ohaus.
Displays Err 9.5
Factory calibration data failed checksum. Perform MFG and MFGWT. If the problem persists reload the TC data. If the
problem still persists the main PCB must be reconfigured at Ohaus.
Displays Err 9.8
User calibration data failed checksum. Perform MFG Calibration. For units with internal calibration also perform MFGWT.
Displays Err 9.9
Can not be fixed in the field the unit has to be temperature compensated.
2.4 EXPLORER PRO ERROR CODE TABLE (Cont.)