SOCT Copernicus PLUS - Service Manual
4. If the height of spectrum does not fit in the correct range arrange it using filter of reference line
Regulation of filter on reference line.
5. If spectrum’s height achieved correct value turn off program skantest.
6. Turn on program Copernicus and perform control scan.
7. If the quality of scan is correct finish service activities.
8. If the quality is incorrect return to the program scantest.
9. Perform points from chapter 13.3 of this instruction.