SOCT Copernicus PLUS - Service Manual
10. Opening and closing the device
10.1 Opening cover
Tools needed to open the cover:
- hexagonal tools size: 2, 2.5, 4, 5,
- magnetic holder, tweezers – just in case when screw is dropped
- Philips-head screwdriver.
1. Read entire manual first before you start opening the SOCT.
2. BE CARREFUL WHEN YOU ARE OPENING THE SOCT PLUS. DO NOT REPLACE POSITION OF
OPTIC COMPONENTS. If you move a position of optics probably you will have to send back
the unit to Optopol Technology Sp. z o.o. for calibration. Especially do not touch the optic
fibre (Red arrows).