SOCT Copernicus PLUS - Service Manual
12.4.2.Exchanging transitional board 151-525.
(For devices from S/N: 154012/I to S/N: 154196/L).
1. Unplug power cord from SOCT device.
2. Open the cover (see: Opening cover).
3. Undo four screws holding transitional board.
4. Disconnect all plugs from transitional board.
5. Take the transitional board out of the device.
6. Fix new transitional board.
7. Do screws holding transitional board.