SOCT Copernicus PLUS - Service Manual
12.5.2 Exchanging board 151-506
Unplug device from power socket .
Open cover (see: Opening cover).
Disconnect all plugs from board 151-506.
Unscrew 4 mounting screws.
Assembly new board.
Plug all plugs according to scheme.