SOCT Copernicus PLUS - Service Manual
130
13.5 Too high spectrum on skantest
1. When the spectrum on skantest is too high after turning on and using function Fast.
2. Measure the power from the objective lane (see: Checking and adjusting power on objective
line). When the power is incorrect arrange it to the normal value.
3. Perform steps from 3 to 9 from chapter 13.4 of this instruction.