DigitalFlow™ DF868 Multipurpose Ultrasonic Liquid Flowmeter Programming Manual (2-Channel) 37
Chapter 1. Programming Site Data
Setting Up Alarms
Complete the following steps to set up the three alarm relays of an option card installed in Slot x:
1. Press [F1]-[F3] to set up alarm relays A, B, or C, respectively.
Note: The set up of alarm A is used as an example. Identical procedures would be used to set up the other alarms.
2. Press [F1]=OFF to disable Alarm A and return to the previous prompt, or press [F2]=HIGH, [F3]=LOW, [F4]=FAULT
to specify Alarm A as the type indicated.
3. Press [F1] for standard operation or [F2] for Failsafe operation. See Chapter 1, Installation, of the Startup Guide for
wiring instructions for each mode.
4. Use the [
 and [ and [F1]-[F4] keys to select a Channel option for the data source, as shown in Table 20 on
page 35.
5. The program now varies, depending on your selection in step 2.
If you selected FAULT as the alarm type, press [F1]-[F3] to select the type of error (flow, non-flow, or both) you
wish to trigger the alarm.
Note: If you have enabled the Energy Option, the DF868 treats the E8 and E9 errors (supply and return
temperature) as flow errors.
If you selected HIGH or LOW:
a.
Use the [ and [ and [F1]-[F4] keys to specify the desired Output Measurement parameter. See Table 21 on
page 36 for a description of the available options.
Note: The measurement units that appear in these prompts are those selected in the GLOBL-SYSTM menu earlier
in this chapter.
b. Enter a value for the trigger point of the alarm and press [ENT].
c. Press [F1]-[F3] to set up another alarm or press [EXIT] to select another slot for set up.
Option Card Totalizer/Frequency Outputs
Complete the following steps to set up the four totalizer/frequency outputs of an option card installed in Slot x:
1. Press [F1]-[F4] to set up outputs A, B, C or D, respectively.
Note: The set up of output A is used as an example. Identical procedures would be used to set up the other
outputs.
2. At the Slot X Pulse Out prompt, press [F1] to disable output A and return to the previous prompt, or press
[F2]=FREQ or [F3]=TTLZR to designate output A as a frequency or a totalizer output, respectively.
3. Use the [
 and [ and [F1]-[F4] keys to select a Channel option for the data source, as shown in Table 20 on
page 35.
Proceed to step 4 on page 38 for instructions on programming a frequency output or to step 9 on page 39 for
instructions on programming a totalizer output. If the selected output was disabled above, skip both sections.