5Patient Alarms and INOPs Te c h n i c a l A l a r m M e s s a g e s ( I N O P s )
72
LA LEAD OFF
Numeric is replaced by -?- for 10
seconds; INOP tone.
ECG The LA electrode has become detached from the patient or the
lead set has been changed. Reattach the electrode or select
New
Lead Setup
in the Setup ECG menu to confirm the
new lead set.
LAP INOPs PRESS See <Pressure label> INOPS (under Pressure).
LEADSET UNPLUGGED Telemetry The leadset has been unplugged from the telemetry device.
LL LEAD OFF
Numeric is replaced by -?- for 10
seconds; INOP tone.
ECG The LL electrode has become detached from the patient or the
lead set has been changed. Reattach the electrode or select
New
Lead Setup
in the Setup ECG menu to confirm the
new lead set.
MCC Reversed
INOP tone
Monitor The MSL coupling cable is reversed. Connect the end with the
grey connector to the Intelligent Display.
MCC Unsupported
INOP tone
Monitor An MSL coupling cable has been connected to a device which
does not support MSL coupling.
Meas. DEACTIVATED An X2 or MP5 has been connected to a host monitor
(companion mode) and all derived measurements have been
deactivated and/or measurements with a label conflict. The
measurements can only be reactivated by disconnecting the
measurement device from the host monitor.
MEASSRV UNSUPPORTD
INOP tone
MMS The Multi-Measurement module is not supported by the
monitor. Contact your service personnel.
MMS Ext. EQUIP MALF
INOP tone
MMS
Extension
Loss of communication between the Multi-Measurement
Module and the MMS extension. Contact your service
personnel.
MMS Ext. UNPLUGGED
INOP tone
MMS
Extension
The MMS extension has been disconnected from the Multi-
Measurement Module.
MMS Ext. Unpowered
INOP tone
MMS
Extension
The MMS extension cannot operate while the Multi-
Measurement Module is running on battery power.
MMSExt.Unsupported
INOP tone
MMS
Extension
The MMS extensions not supported by your monitor. Contact
your service personnel.
MMS UNPLUGGED
INOP tone.
MMS Make sure that the Multi-Measurement Module is connected
to the monitor. All MMS measurements are off while the MMS
is unplugged.
MMS UNSUPPORTED
INOP tone.
MMS The Multi-measurement Module is not supported by your
monitor. Contact your service personnel.
!!/!!!MORE BED ALARMS
At Information Center
Monitor The monitor is associated with a telemetry device and is
sending data to the Information Center via the telemetry
device. There are currently more alarms at the bedside than can
be transmitted to the Information Center.
MSL Power High Monitor The power consumption of the devices connected to the
Measurement Link (MSL) cable is too high. If this situation
continues, the MSL will be switched off. Contact your service
personnel.
MSL Power Off
INOP tone.
Monitor The power consumption of the devices connected to the
Measurement Link (MSL) cable was too high for too long and
the MSL has been switched off. Contact your service personnel.
MSL Power Overload
INOP tone.
Monitor The power consumption of the devices connected to the
Measurement Link (MSL) cable is much too high or there has
been a short circuit. The MSL has been switched off. Contact
your service personnel.
INOP Message, Indication Source What to do