9.8 Calibrating the Standard Signal
In the restricted operating menu: The
DULCOMETER
®
D1Cb / D1Cc calibrates the zero
point.
In the complete operating menu: The DULCOMETER
®
D1Cb / D1Cc carries out a two-
point calibration.
WARNING!
Incorrect metering due to incorrect metering range
Possible consequence: Fatal or serious injuries
– The measuring range of the sensor or the measuring device is essential for the
measuring range!
– If the assignment of the measuring range is modified, the settings must be
checked in all menus
– If the assignment of the measuring range is changed, the sensor or the meas‐
uring device must be recalibrated
CAUTION!
Correct sensor operation / Run-in time
Damage to the product or its surroundings
– Correct measuring and dosing is only possible if the sensor is working perfectly
–
Please read the operating manual for the sensor or the measuring device
– It is imperative that the run in periods of the sensors or the measuring device are
adhered to
– The run in periods should be allowed for when planning commissioning
– It may take a whole working day to run-in the sensor or the measuring device
Measured Variables and Operating Menus for the Standard Signal General