7.1 Reduced / Complete Operating Menu
The DULCOMETER
®
D1Cb / D1Cc allows settings to be made in two different compre‐
hensive menus (reduced / complete). All of the parameters of the controller are preset
and can be changed in the complete operating menu.
The controller is delivered with a restricted operating menu. If adjustments are neces‐
sary, the parameters can be changed by switching to the complete operating menu.
general setting
information
Identcode:
D1CB xxxxxxxxxx
Vxxxxxxxxxx
Srnr: xxxxxxxxxxx
alarm relay
access c.:
5000
operating menu
english
reduced
A0088_GB
operating menu
Calib timer
wash timer
_
_
Fig. 35: Reduced / complete changeover
7.2 Description of All Amperometric Measured Variables
WARNING!
Danger of incorrect metering
This can result in hazardous incorrect metering
During initial commissioning, the measured variable and the measuring range of the
sensor must be set prior to calibration. Refer to
Ä Chapter 5.1.2 ‘Selection of the
Measured Variable and Measuring Range’ on page 54
Measured variable Default measuring range
Chlorine, chlorine dioxide, ozone 2 ppm
Bromine 10 ppm
Oxygen 20 ppm
Peracetic acid 2000 ppm
Hydrogen peroxide 200 ppm
Chlorite 0.5 ppm
Measured Variables and Operating Menus for Amperometric Sensors