5.1.2 Selection of the Measured Variable and Measuring Range
WARNING!
Incorrect metering due to incorrect measuring range
Possible consequence: Fatal or serious injuries.
– The measuring range of the sensor is essential for the measuring range!
–
If the assignment of the measuring range is modified, the settings must be
checked in all menus
– If the assignment of the measuring range is changed, the sensor must be recali‐
brated
With devices, which have not been preconfigured to the customer's specific requirement,
the measured variable required has to be selected in the complete operating menu "Gen‐
eral Settings / Change Measured Variable". The DULCOMETER
®
D1Cb / D1Cc then has
to be labelled with the label corresponding to the measured variable selected. The rele‐
vant labels are enclosed with the DULCOMETER
®
D1Cb / D1Cc.
general setting
Information
Identcode:
Vxxxxxxxxxxxx
D1CBxxxxxxxxxxx
Srnr: xxxxxxxxxx
add. functions
Software version
measure change
wash timer chlorit
Br
O3
O2
effective range
A0032
0...2.00ppm
Fig. 24: Selection of measured variable and measuring range
The measuring range required has to be selected and set in the complete operating
menu General Settings / Change Measured Variable, see
Ä Chapter 10.6.1 ‘Setting the
Measured Variable/Measuring Range’ on page 131
.
Commissioning
54