only with correction value
correction value
setting ?
correcting value
temperature
automatic
A0020_GB
Fig. 40: Correcting value
The correction variable compensates for the effect of the temperature of the medium on
the measured value. The correction variable is the temperature of the medium to be
measured. The temperature of the medium affects the value to be measured. For
amperometric sensors only necessary when using the DULCOTEST
®
CDP sensor for
chlorine dioxide ClO
2
.
Operating modes
n Off: No temperature compensation takes place.
For measurements which do not require temperature compensation.
n Automatic: The DULCOMETER
®
D1Cb / D1Cc evaluates the temperature signal of
the connected temperature sensor.
For measurements with temperature sensors, which deliver a temperature signal
which can be evaluated by the DULCOMETER
®
D1Cb / D1Cc (Pt100/Pt1000) (0
-100 °C).
n Manual: The temperature of the medium to be measured has to be measured by the
user. The measured value is then entered using the keys:
, and into the
DULCOMETER
®
D1Cb / D1Cc and saved by pressing the key .
For measurements where the medium to be measured has a constant tempera‐
ture, which has to be taken into account in the control process.
Measured Variables and Operating Menus for Amperometric Sensors