8.5.4.2 Testing Redox Sensors Buffer Values Tables
Table: Possible buffer values
Possible values
Setting Starting
value
Increment Lower value Upper value Remarks
Buffer
values
Reading 1 mV -1,500 mV +1,500 mV
185-265 mV 220 mV
425-505 mV 465 mV
8.5.4.3 Testing Redox Sensors. Description of the Error Messages
Table: Possible error messages when testing redox sensors
Error message Condition Effect
Reading high Reading
40 mV> buffer
Back to the permanent dis‐
play
Basic load metering
Reading low Reading
40 mV< buffer
Back to the permanent dis‐
play
Basic load metering
Measured Variables and Operating Menus for Potentiometric Sensors
94