Chapter 2 Command System RIGOL
DP800 Programming Guide 2-13
IEEE488.2 Common Commands
Command List:
*IDN?
*RCL
*RST
*SAV
*TST?
*IDN?
Syntax *IDN?
Description Query the ID character string of the instrument.
Return Format The query returns the ID character string of the instrument, for example, RIGOL
TECHNOLOGIES,DP831A,DP8A000001,00.01.01.
*RCL
Syntax *RCL {1|2|3|4|5|6|7|8|9|10}
Description Read the instrument state stored.
Parameter
Name Type Range Default
{1|2|3|4|5|6|7|8|9|10} Discrete 1|2|3|4|5|6|7|8|9|10 None
Explanation The power supply provides 10 storage locations (numbered 1 to 10) for
instrument states. This command reads the instrument state stored in the
specified location.
This command is only available when the specified storage location contains a
state file.
Related
Command
:MEMory[:STATe]:LOAD
*RST
Syntax *RST
Description Restore the power supply to factory state (refer to
Appendix B: Factory Setting) and
clear the error queue.
Related
Commands
:PRESet:KEY
:PRESet[:APPLy]

Table of Contents

Other manuals for Rigol DP832

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DP832 and is the answer not in the manual?

Rigol DP832 Specifications

General IconGeneral
Number of Outputs3
Output Voltage Range CH1/CH20-30V
Output Voltage Range CH30-5V
Output Current Range CH1/CH20-3A
Output Current Range CH30-3A
Total Power195 W
Power per Channel CH1/CH290 W
Power per Channel CH315 W
Voltage Resolution1 mV
Current Resolution1 mA
InterfacesUSB, RS232
Display3.5 inch TFT LCD