EasyManuals Logo

Rigol DP832 Programming Guide

Rigol DP832
134 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #76 background imageLoading...
Page #76 background image
RIGOL Chapter 2 Command System
2-58 DP800 Programming Guide
:SYSTem:COMMunicate:RS232:DATABit
Syntax :SYSTem:COMMunicate:RS232:DATABit {5|6|7|8}
:SYSTem:COMMunicate:RS232:DATABit?
Description Set the data bit of the RS232 interface.
Query the data bit of the RS232 interface.
Parameter
Name Type Range Default
{5|6|7|8} Discrete 5|6|7|8 8
Return Format The query returns 5, 6, 7 or 8.
Example :SYST:COMM:RS232:DATAB 8
:SYST:COMM:RS232:DATAB?
:SYSTem:COMMunicate:RS232:FLOWCrl
Syntax :SYSTem:COMMunicate:RS232:FLOWCrl {ON|OFF}
:SYSTem:COMMunicate:RS232:FLOWCrl?
Description Enable or disable the hardware flow control.
Query the status of the hardware flow control.
Parameter
Name Type Range Default
{ON|OFF} Bool ON|OFF OFF
Explanation This power supply uses RTS/CTS hardware flow control mode. The instrument
monitors the status of the CTS pin. When the status is True, the instrument sends
data; when the status is “False, the instrument stops sending data. The instrument
sets the CTS pin to “False” when the input buffer area is almost full and sets the CTS
pin to Truewhen the input buffer area is available again.
Return Format The query returns ON or OFF.
Example :SYST:COMM:RS232:FLOWC ON
:SYST:COMM:RS232:FLOWC?
:SYSTem:COMMunicate:RS232:PARItybit
Syntax :SYSTem:COMMunicate:RS232:PARItybit {NONE|ODD|EVEN}
:SYSTem:COMMunicate:RS232:PARItybit?
Description Set the parity mode to None, Oddor Even.
Query the current parity mode.
Parameter
Name Type Range Default
{NONE|ODD|EVEN} Discrete NONE|ODD|EVEN NONE
Return Format The query returns NONE, ODD or EVEN.
Example :SYST:COMM:RS232:PARI ODD
:SYST:COMM:RS232:PARI?

Table of Contents

Other manuals for Rigol DP832

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DP832 and is the answer not in the manual?

Rigol DP832 Specifications

General IconGeneral
Number of Outputs3
Output Voltage Range CH1/CH20-30V
Output Voltage Range CH30-5V
Output Current Range CH1/CH20-3A
Output Current Range CH30-3A
Total Power195 W
Power per Channel CH1/CH290 W
Power per Channel CH315 W
Voltage Resolution1 mV
Current Resolution1 mA
InterfacesUSB, RS232
Display3.5 inch TFT LCD