EasyManuals Logo

Rigol DP832 Programming Guide

Rigol DP832
134 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #57 background imageLoading...
Page #57 background image
Chapter 2 Command System RIGOL
DP800 Programming Guide 2-39
:PRESet:USER[n]:SET:CURRent
Syntax :PRESet:USER[n]:SET:CURRent {<current>|MINimum|MAXimum}
:PRESet:USER[n]:SET:CURRent? [MINimum|MAXimum]
Description Set the current of the specified user-preset state.
Query the current of the specified user-preset state.
Parameter
Name Type Range Default
[n] Discrete 1|2|3|4 None
<current> Real Current range of the current channel
Refer to the
Explanation
Explanation By default, this command sets the parameter of the channel currently selected.
If you want to set the parameter of other channel, please use
the
:INSTrument[:SELEct] command to select the desired channel.
The current default values of each user-preset state are as shown in the table
below.
DP831A DP832A/DP832
CH1 CH2 CH3 CH1 CH2 CH3
USER1 0.5000A 1.0000A 1.0000A 1.0000A 1.0000A 1.0000A
USER2 1.0000A 1.0000A 1.0000A 1.0000A 1.0000A 1.0000A
USER3 1.0000A 1.5000A 1.5000A 2.0000A 2.0000A 1.0000A
USER4 1.0000A 2.0000A 2.0000A 3.0000A 3.0000A 2.0000A
Return Format The query returns the current, for example, 1.5000.
Example :PRES:USER1:SET:CURR 1.5
:PRES:USER1:SET:CURR?
:PRESet:USER[n]:SET:DEFault
Syntax :PRESet:USER[n]:SET:DEFault
Description Restore the parameters of the specified user-preset state to default values.
Parameter
Name Type Range Default
[n] Discrete 1|2|3|4 None
Explanation For the default values of the user-preset states, refer to the Explanationof
the
:PRESet:USER[n]:SET:CURRent and :PRESet:USER[n]:SET:VOLTage
commands.
Related
Commands
:PRESet:USER[n]:SET:CURRent
:PRESet:USER[n]:SET:VOLTage

Table of Contents

Other manuals for Rigol DP832

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DP832 and is the answer not in the manual?

Rigol DP832 Specifications

General IconGeneral
Number of Outputs3
Output Voltage Range CH1/CH20-30V
Output Voltage Range CH30-5V
Output Current Range CH1/CH20-3A
Output Current Range CH30-3A
Total Power195 W
Power per Channel CH1/CH290 W
Power per Channel CH315 W
Voltage Resolution1 mV
Current Resolution1 mA
InterfacesUSB, RS232
Display3.5 inch TFT LCD

Related product manuals