RIGOL Chapter 2 Command System
2-54 DP800 Programming Guide
:SYSTem:COMMunicate:LAN:DHCP[:STATe]
Syntax
:SYSTem:COMMunicate:LAN:DHCP[:STATe] {ON|OFF}
:SYSTem:COMMunicate:LAN:DHCP[:STATe]?
Description Enable or disable the DHCP mode.
Query the status of the DHCP mode.
Parameter
Name Type Range Default
{ON|OFF} Bool ON|OFF ON
Explanation In DHCP mode, the DHCP server in the current network assigns network
parameters (such as the IP address) for the instrument.
When all the three configuration modes are set to On, the priority order of
parameter configuration is DHCP, AutoIPand “ManualIP”.
The three IP configuration modes cannot all be set to Offat the same time.
Return Format The query returns ON or OFF.
Example
:SYST:COMM:LAN:DHCP ON
:SYST:COMM:LAN:DHCP?
Related
Commands
:SYSTem:COMMunicate:LAN:AUTOip[:STATe]
:SYSTem:COMMunicate:LAN:MANualip[:STATe]
:SYSTem:COMMunicate:LAN:DNS
Syntax :SYSTem:COMMunicate:LAN:DNS <dns>
:SYSTem:COMMunicate:LAN:DNS?
Description Set the DNS (Domain Name Service).
Query the current DNS.
Parameter
Name Type Range Default
<dns>
ASCII character
string
The format is nnn.nnn.nnn.nnn;
the first nnn ranges from 1 to 223
(except 127), the other three range
from 0 to 255
None
Explanation This command is only available when the manual IP configuration mode is
enabled.
You are recommended to ask your network administrator for an address
available.
Return Format The query returns the DNS address, for example, 172.16.3.2.
Example :SYST:COMM:LAN:DNS 172.16.3.2
:SYST:COMM:LAN:DNS?
Related
Command
:SYSTem:COMMunicate:LAN:MANualip[:STATe]

Table of Contents

Other manuals for Rigol DP832

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DP832 and is the answer not in the manual?

Rigol DP832 Specifications

General IconGeneral
Number of Outputs3
Output Voltage Range CH1/CH20-30V
Output Voltage Range CH30-5V
Output Current Range CH1/CH20-3A
Output Current Range CH30-3A
Total Power195 W
Power per Channel CH1/CH290 W
Power per Channel CH315 W
Voltage Resolution1 mV
Current Resolution1 mA
InterfacesUSB, RS232
Display3.5 inch TFT LCD