Chapter 2 Command System RIGOL
DP800 Programming Guide 2-65
:TIMEr:ENDState
Syntax :TIMEr:ENDState {OFF|LAST}
:TIMEr:ENDState?
Description Set the end state of the timer to Offor Last.
Query the current end state of the timer.
Parameter
Name Type Range Default
{OFF|LAST} Discrete OFF|LAST OFF
Explanation The end state refers to the state of the instrument after it finishes outputting
the total number of groups of voltage/current values when the number of
cycles is a specified value.
OFF: the instrument turns off the output automatically after finishing the
output.
LAST: the instrument stops at the output state of the last group after finishing
the output.
The total number of groups in each timing output = the number of groups ×
the number of cycles; wherein, the number of groups is set by
the
:TIMEr:GROUPs command and the number of cycles is set by
the
:TIMEr:CYCLEs command.
Return Format The query returns OFF or LAST.
Example :TIME:ENDS LAST
:TIME:ENDS?
Related
Commands
:TIMEr:CYCLEs
:TIMEr:GROUPs

Table of Contents

Other manuals for Rigol DP832

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DP832 and is the answer not in the manual?

Rigol DP832 Specifications

General IconGeneral
Number of Outputs3
Output Voltage Range CH1/CH20-30V
Output Voltage Range CH30-5V
Output Current Range CH1/CH20-3A
Output Current Range CH30-3A
Total Power195 W
Power per Channel CH1/CH290 W
Power per Channel CH315 W
Voltage Resolution1 mV
Current Resolution1 mA
InterfacesUSB, RS232
Display3.5 inch TFT LCD

Related product manuals