RIGOL Chapter 2 Command System
2-28 DP800 Programming Guide
:MONItor:STOPway
Syntax :MONItor:STOPway {OUTOFF|WARN|BEEPER|NONE},{ON|OFF}
:MONItor:STOPway?
Description Set the stop mode of the monitor. Multiple stop modes can be enabled at the same
time.
Query the stop mode of the monitor.
Parameter
Name Type Range Default
{OUTOFF|WARN|BEEPER|
NONE}
Discrete OUTOFF|WARN|BEEPER None
{ON|OFF} Bool ON|OFF ON
Explanation When the channel output state meets the monitor condition set, the instrument will
turn off the output, display the corresponding prompt message or sound the beeper
according to the stop mode selected.
Return Format The query returns the on/off status of OUTOFF, WARN and BEEPER, for example,
OutputOff:ON,Warn:ON,Beep:ON.
Example :MONI:STOP BEEPER,OFF /*Disable the Beeperstop mode*/
:MONI:STOP? /*Query the current states of the stop modes of the monitor*/
/*The query returns OutputOff:ON,Warn:ON,Beep:OFF*/

Table of Contents

Other manuals for Rigol DP832

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DP832 and is the answer not in the manual?

Rigol DP832 Specifications

General IconGeneral
Number of Outputs3
Output Voltage Range CH1/CH20-30V
Output Voltage Range CH30-5V
Output Current Range CH1/CH20-3A
Output Current Range CH30-3A
Total Power195 W
Power per Channel CH1/CH290 W
Power per Channel CH315 W
Voltage Resolution1 mV
Current Resolution1 mA
InterfacesUSB, RS232