Chapter 11 Protocol Decoding    RIGOL 
MSO7000/DS7000 User Guide  11-33 
⚫  CRC (Cyclic Redundancy Check): expressed in Hex, displayed in 
yellow-green. When errors occur, it is displayed as a red patch. 
⚫  ACK (Acknowledgement): when effective, displayed as a light yellow-green 
patch;  when  errors  (ACK  is  detected  to  be  1)  occur,  displayed  as  a  red 
patch.   
 
 
 
 
 
 
⚫  R (remote frame): displayed as a tangerine patch. 
 
 
 
FlexRay Decoding (Option) 
 
FlexRay is a type of differential serial bus configured with three consecutive 
segments (i.g. header, payload, and trailer). The oscilloscope samples the FlexRay 
signal at the specified sample position and judges each data point as logic "1" or 
logic "0" according to the preset threshold level. The FlexRay decoding is required to 
specify the signal type and baud rate. 
 
In the decode setting menu, press Decode1 → Bus Type, then rotate the 
multifunction knob    to select "FlexRay". Press down the knob to select it. You can 
also press Bus Type continuously or enable the touch screen to select it. 
 
1.  Enable or disable the bus 
Press Bus Status continuously to enable or disable the decoding function. 
 
2.  Quickly apply FlexRay trigger settings to FlexRay decoding 
Press Copy Trig to copy the FlexRay trigger settings and apply them to FlexRay 
decoding  function  (set  the  corresponding  FlexRay  decoding  parameters 
automatically).  For  analog  channels,  the  threshold  level  settings  will  also  be 
copied. 
 
3.  Set the source and the threshold 
⚫  Press Source and rotate the multifunction knob    to select the desired 
source channel, and then press down the knob to select it. You can also 
press Source continuously or enable the touch screen to select it. The 
available channels include CH1-CH4 and D0-D15.