Chapter 1 Quick Start    RIGOL 
MSO7000/DS7000 User Guide  1-9 
Function Inspection 
 
1.  Press Default on the front panel, then a prompt message "Restore default?" is 
displayed. Press OK or tap OK to restore the instrument to its factory default 
settings. 
2.  Connect  the  ground  alligator  clip  of  the  probe  to  the  "Ground  Terminal"  as 
shown in Figure 1-10 below. 
3.  Use the probe to connect the input terminal of CH1 of the oscilloscope and the 
"Compensation Signal Output Terminal" of the probe, as shown in Figure 1-10. 
 
 
 
Figure 1-10 To Use the Compensation Signal 
 
4.  Set the probe attenuation to 10X, and then press AUTO. 
5.  Observe the waveform on the display. In normal condition, the square waveform 
as shown in Figure 1-11 should be displayed. 
 
Figure 1-11 Square Waveform Signal 
 
6.  Use  the  same  method  to  test  the  other  channels.  If  the  square  waveforms 
actually shown do not match that in the figure above, please perform "Probe 
Compensation" introduced in the next section. 
 
WARNING 
To avoid electric shock when using the probe, please make sure that the 
insulated  wire  of  the  probe  is  in  good  condition.  Do  not  touch  the 
metallic part of the probe when the probe is connected to high voltage 
source.