Chapter 4 To Set the Sample System    RIGOL     
MSO7000/DS7000 User Guide  4-11 
LA Memory Depth 
 
Press Acquire → More to view the memory depth of the current digital channel 
under LA Mem Depth. The LA memory depth will change with the memory depth of 
the analog channel, and cannot be set separately. The maximum LA memory depth is 
62.5 Mpts. 
 
 
Anti-Aliasing 
 
At a slower sweep speed, the sample rate is reduced, and a dedicated display 
algorithm can be used to minimize the possibility of aliasing. 
 
Press Acquire → Anti-aliasing to enable or disable the anti-aliasing function. By 
default, anti-aliasing is disabled. In this case, it is more likely to generate waveform 
aliasing. 
 
Note: This function should be enabled when the horizontal time base has a large 
deviation from the period of the signal under test (e.g. when the horizontal 
time base is greater than five-fold the period of the signal under test). 
 
 
Horizontal Expansion 
 
Horizontal expansion indicates the reference position that the screen waveform is 
referenced to when it is horizontally expanded or compressed in adjusting the 
Horizontal    SCALE knob. In YT mode, press Acquire → More → Expand and 
then rotate the multifunction knob    to select the desired reference position. Press 
down the knob to select it. You can also press Expand continuously to select it or 
enable the touch screen to tap the desired reference. The horizontal expansion 
reference supported by the oscilloscope includes Center, Left, Right, Trigger, and User. 
The default is "Center". 
⚫  Center:  when  the  horizontal  time  base  is  modified,  the  waveform  will  be 
expanded or compressed horizontally relative to the screen center. 
⚫  Left: when the horizontal time base is modified, the waveform will be expanded 
or compressed horizontally relative to the leftmost position of the screen. 
⚫  Right:  when  the  horizontal  time  base  is  modified,  the  waveform  will  be 
expanded or compressed horizontally relative to the rightmost position of the 
screen. 
⚫  Trigger:  when  the  horizontal  time  base  is  modified,  the  waveform  will  be 
expanded or compressed horizontally relative to the trigger point. 
⚫  User: when the horizontal time base is modified, the waveform displayed will be 
expanded  or  compressed  horizontally  relative  to  the  user-defined  reference 
position.