RIGOL Chapter 1 Quick Start
1-14 MSO7000/DS7000 User Guide
6. EXT TRIG
Inputs the external signal via the external input terminal.
7. Trigger Out and Pass/Fail
âš« TRIG OUT:
The oscilloscope can output a signal that can reflect the current
capture rate of the oscilloscope at each trigger via this interface.
Connect the signal to a waveform display device and measure the
frequency of the signal. The measurement result is the same as the
current capture rate.
âš« Pass/Fail
The instrument can output a pulse from the [TRIG OUT] connector
when a pass/failed event is detected during the pass/fail test.
8. Lock Hole
You can lock the instrument to a fixed location by using the security lock
(please purchase it by yourself) via the lock hole.
9. Fuse
If you need to replace the fuse, use only the specified fuse. For details, refer
to descriptions in "To Replace the Fuse".
10. AC Power Cord Connector
Indicates the input terminal where AC power source is connected. The
power supply requirements of the instrument are: 100 V~240 V; 45
Hz~440 Hz. Please use the power cord provided in the accessories to
connect the oscilloscope to the AC power source.

Table of Contents

Other manuals for Rigol DS7024

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS7024 and is the answer not in the manual?

Rigol DS7024 Specifications

General IconGeneral
BrandRigol
ModelDS7024
CategoryTest Equipment
LanguageEnglish

Related product manuals