138
Serial Interface
Data output
DOD
Output displayed value
Send the request command at one second interval or longer.
Request command DOD?
Response data d1,d2,...,d67
Returned value d1 = “xxx.x” Acc. of X channel (Inst value)
d2 = “xxx.x” PPV of X channel (Inst value)
d3 = “xxx.x” D.F. of X channel (Inst value)
d4 = “xx.xx” Disp. of X channel (Inst value)
d5= 0 or 1 Inst Overload information
(1: Yes, 0: No)
d6= 0 or 1 Inst Under-range information
(1: Yes, 0: No)
d7 = “xxx.x” Acc. of X channel (Max hold
value)
d8 = “xxx.x” PPV of X channel (Max hold
value)
d9 = “xxx.x” D.F. of X channel (Max hold
value)
d10 = “xxx.x” Disp. of X channel (Max hold
value)
d11 = 0 or 1 Max hold Overload information
(1: Yes, 0: No)
d12 = 0 or 1 Max hold Under-range information
(1: Yes, 0: No)
d13 = “xxx.x” PPV of X channel (Calc value)
d14 = “xxx.x” D.F. of X channel (Calc value)
d15 = “xxx.x” Disp. of X channel (Calc value)
d16 = “xxx.x” Acc. of X channel (Calc value)
d17 = “xxx.x” C.F. of X channel (Calc value)
d18 = “xxx.x” MTVVof X channel (Calc value)
d19 = “xxx.x” VDV of X channel (Calc value)