146
Reference Information
INPUT connector
The input connector is wired as shown below.
A: Unused
B: Ground
C: Z channel signal input
D: Unused
E: X channel signal input
F: Y channel signal input
G: Pickup power supply
Important
Do not connect anything else besides the
supplied Vibration Pickup PV-83D. Otherwise
damage may occur.
Calibration example using an exciter
The following arrangement can be used to calibrate the vibration pickup with an
exciter.
• Sine wave signal oscillator
• Power ampli er DC-300A series II (Manufacturer:Crown)
• Exciter VG-100 (Vertical-directionexciter)
(Manufacturer:AR Brown)
APS-129 (Horizontal-direction exciter)
(Manufacturer:AR Brown)
• Reference accelerometer PV-03 (Manufacturer:RION)
• Vibration meter VM-83 (Manufacturer:RION)
A
B
C
D
E
F
G
Top view