EasyManua.ls Logo

Rion VM-56 - Calibration Example Using an Exciter; INPUT Connector

Rion VM-56
180 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
146
Reference Information
INPUT connector
The input connector is wired as shown below.
A: Unused
B: Ground
C: Z channel signal input
D: Unused
E: X channel signal input
F: Y channel signal input
G: Pickup power supply
Important
Do not connect anything else besides the
supplied Vibration Pickup PV-83D. Otherwise
damage may occur.
Calibration example using an exciter
The following arrangement can be used to calibrate the vibration pickup with an
exciter.
• Sine wave signal oscillator
• Power ampli er DC-300A series II (Manufacturer:Crown)
• Exciter VG-100 (Vertical-directionexciter)
(Manufacturer:AR Brown)
APS-129 (Horizontal-direction exciter)
(Manufacturer:AR Brown)
• Reference accelerometer PV-03 (Manufacturer:RION)
• Vibration meter VM-83 (Manufacturer:RION)
A
B
C
D
E
F
G
Top view