o If the Na
+
intercept (b) and the slope (m) are correct, press YES.
o To change these values, press NO. The intercept and slope values may be adjusted
by pressing NO to change the number, and by pressing YES to accept the
number. The programmed factors are automatically applied to blood/serum and
QC samples.
Follow the same procedure for the intercept (b) and slope (m) for K
+
, Cl
-
, Ca
2+
and
Li
+
.
q
Only activated parameters can be programmed.
Separate correlation factors are available for urine samples.
o At the prompt Program Urine Corr Factors?, press YES.
o Program the correlation factors as described above.
q
There are no urine correlation factors available for Ca
2+
and Li
+
. These parameters are not
available for measurement in urine mode.