124 Instructions for Use · Version 10.0
8 Quality control 9180 Electrolyte Analyzer
Performing a QC measurement
o The prompt Store values in memory? appears.
o To save the values in memory, press YES. To reject the values, press NO.
o If the values are rejected, the analyzer will return to the QC Level 1 Sample?
display, allowing to repeat the QC Level 1 measurement by pressing YES, or to
skip to QC Level 2
q
In case the value is outside the measurement range or ERR, it will automatically be
rejected.
If the values have been saved, Values Accepted! will be displayed shortly, followed by
the prompt QC Level 2 Sample?.
For further QC measurements (Level 2 and Level 3), follow the outlined instructions
for Level 1, being sure to use a Level 2 or Level 3 ISETROL
To end the QC measurement, press the NO key until READY is displayed.
After finishing the QC Level 3 measurement, the analyzer displays
Remain in QC/Std./Urine Sample?. If measurements for all levels have been
completed, press NO and the analyzer returns to the READY display. If you press
YES, the analyzer returns to QC Level 1 Sample?.