Roche Diagnostics
134 Instructions for Use · Version 10.0
9 Maintenance 9180 Electrolyte Analyzer
Daily maintenance
o Hold the Cleaning Solution under the probe until
Wipe Probe/Close Sample Door is displayed. Then, use a lint-free tissue to
remove the cleaning solution from the probe. Close the sample door.
Figure 9-1
o While the analyzer displays Thank You! and a brief countdown, pour a small
amount of Sodium Electrode Conditioner into a clean container.
o At the prompt Perform Daily Conditioning?, press YES.
o At the prompt Open Sample Door Introduce Sample, lift the sample door and
the pump will begin to aspirate.
o Hold the Sodium Electrode Conditioner under the probe until
Wipe Probe/Close Sample Door is displayed. Then, use a lint-free tissue to
remove the Sodium Electrode Conditioner from the probe. Close the sample
door.
o After the analyzer displays Thank You! and a brief countdown, press NO at the
prompt Remain in Daily Maintenance?. An automatic calibration will be started.