Roche Diagnostics
90 Instructions for Use · Version 10.0
6 Installation and shutdown 9180 Electrolyte Analyzer
Installation
o Hold the Cleaning Solution under the probe until
Wipe Probe/Close Sample Door is displayed. Use a lint-free tissue to remove the
cleaning solution from the probe. Close the sample door.
Figure 6-17
o While the analyzer displays Thank You! and a brief countdown, pour a small
amount of Sodium Electrode Conditioner into a clean container
o Answer the prompt Perform Daily Conditioning? and press YES.
o At the prompt Open Sample Door Introduce Sample, lift the sample door. The
pump will begin to aspirate.
Figure 6-18
o Hold the Sodium Electrode Conditioner under the probe until
Wipe Probe/Close Sample Door is displayed. Use a lint-free tissue to remove the
Sodium Electrode Conditioner from the probe. Close the sample door.
Figure 6-19