SECTION IV
MAINTENANCE / SERVICE
Page 4-67Printed in U.S.A.
(5) If status switch (4) needs to be adjusted,
remove switch bracket (2) and adjust status
switch as determined necessary. Reinstall
switch bracket.
(6) Raise table to its upright position.
(7) Using hand control, run UNLOCK function to
unlock floor lock cylinders and then run LOCK
function to lock floor lock cylinders.
(8) Observe. When floor lock cylinders are un-
locked, hand control should read ON WHEELS.
When floor lock cylinders are locked, hand
control should read FLOOR LOCKED. Error
code E01 (Floor Lock Status Switch Is Not
Responding) should not display.
Figure 4-71. Floor Lock Status Switch
Removal / Installation
BOTTOM VIEW
LOCK
POSITION
UNLOCK
POSITION
1
2
3
4
5
7
6
MAIN FLOOR LOCK
STATUS SWITCH
OUTRIGGER FLOOR LOCK
STATUS SWITCH
HA551100
© Schaerer Mayfield USA, Inc. 2004