SECTION II
TESTING AND TROUBLESHOOTING
Table 2-1. Troubleshooting Guide
Page 2-16Printed in U.S.A.
Problem Symptom Probable Cause Check Correction
Seat section moves
improperly during return to
level function.
Seat section does not
return to proper position
when the "Level" button is
pressed.
Screws securing seat
position sensor are loose.
Check for loose screws. Tighten loose screws (if
necessary coat threads of
screws with removable
threadlocking adhesive
(Loctite 242). Refer to para
4.19.
© Schaerer Mayfield USA, Inc. 2004