SECTION IV
MAINTENANCE / SERVICE
(7) Using hand control, press TABLE DOWN button
to lower column assembly 1/32 to 1/16 in. (0.8
to 1.6 mm).
(8) With help of an assistant, tilt table over far
enough to access valve (1).
(9) Using a socket and extension, close valve (1).
Tighten valve (1) to 11-12.5 ft-lbs (15-17 N•m).
(10) Using hand control, press TABLE UP and
TABLE DOWN buttons to run the column
function all the way up and down. Check that
column assembly does not jump downward
when column assembly is lowered from a fully
extended position. Check that maximum table
height has not been compromised (Refer to
Table 1-2 for specifications).
Figure 4-7. Column Cylinder Slave Line Volume Adjustment
Page 4-8Printed in U.S.A.
CA806300
HEAD END
SMI
7300
COLUMN
ASSEMBLY
NORMAL
POSITION
REVERSE
POSITION
TREND
REV TREND
HEIGHT UP
HEIGHT DOWN
TILT L
TILT R
SEAT UP SEAT DOWN
FLEX
REFLEX
ENABLE
DISABLE
LOCK
UNLOCKLEVEL
SMI
7300
TABLE
DOWN
TABLE
UP
© Schaerer Mayfield USA, Inc. 2004