94seca 655/654
17-10-07-654-100_01-2020B
4. Tighten the knurled wheels in the direction of the arrow.
The foot screws are secured against being adjusted.
5. Calibrate the ultrasonic measuring rod Calibrating ultrasonic height
measurement, page 109.
5.4 Connecting a barcode scanner (optional)
A barcode scanner can be connected to the USB interface of the weighing
platform.
The barcode scanner is required for the following functions:
Configuration: Define network data in the
seca connect 103 software
and transfer them to the device using the QR code:
Setting up
network functions, page 115
Operation: Record patient and user IDs for transmitting measuring re-
sults to an EMR System: Completing the measurement procedure,
page 105
WARNING!
Patient hazard
Route the connecting cable so that patients cannot become caught
or strangle themselves in it.
Route the connecting cable so that patients and users cannot
stumble.
NOTE
Observe the maximum permitted current consumption of the
barcode scanner General technical data, page 135.
Use only barcode scanners recommended by seca Optional
accessories and spare parts, page 137.
To connect a barcode scanner, proceed as follows:
1. Make sure that the device is disconnected from the power supply.
2. Plug the USB connector of the barcode scanner into the USB socket of
the multifunctional display.
3. Hang the barcode scanner in a suitable holder (e.g. scanner holder
seca 463 Optional accessories and spare parts, page 137.