EasyManua.ls Logo

Seca 654 - 6.4 Measuring weight and height (devices with ultrasonic measuring rod)

Seca 654
67 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation • 99
English
17-10-07-654-100_01-2020B
6.4 Measuring weight and height (devices with ultrasonic measuring rod)
CAUTION!
Patients can be injured if they fall
Persons with limited mobility can fall when stepping onto the weighing
platform.
Support people with limited mobility when they step onto the scale.
WARNING!
Incorrect measurement due to reflections
If there are objects or people in the immediate vicinity of the device,
incorrect measurements will result.
Ensure that there are no objects or people within 0.5 meter of the
front or side of the device during the measuring process.
Ensure that the device is at least 0.2 meter away from the wall.
Ensure that the patient is not wearing any kind of hair accessory on
top of the head.
The measurement procedure described on the following is based on the
factory settings. Information about configuration options can be found here:
Configuring voice guidance, page 119.
1. Make sure that there is no load on the weighing platform.
2. If necessary, press the multifunctional display screen to “wake the device
from standby.
3. Make sure that the Status LED on the ultrasound head and the silhouettes
on the weighing platform light up.
4. Make sure that the patient steps forward onto the weighing platform and
adopts an upright posture.
5. Make sure that the patient stands on the illuminated foot silhouettes.
6. Ask the patient to follow the instructions given by the device.
The device determines the patient's weight and height.
Device mode Function available
Basic
Advanced
Expert
Service
Foot silhouettes
Status LED

Table of Contents

Other manuals for Seca 654

Related product manuals