Technical data • 135
17-10-07-654-100_01-2020B
12.3 General technical data
General technical data
Ambient conditions, operation
• Temperature
• Air pressure
•Humidity
+10 °C to +40 °C (50 °F to 104 °F)
700 hPa – 1060 hPa
20 % – 80 %, noncondensing
Ambient conditions, storage
• Temperature
• Air pressure
•Humidity
• Warm-up time from lowest storage temperature to operational
temperature
- At ambient temperature 20 °C:
- At ambient temperature 20 °C and with condensation formation:
• Cooling time from highest storage temperature to operational
temperature (at ambient temperature 20 °C)
-10 °C to +65 °C (14 °F to 149 °F)
700 hPa – 1060 hPa
0 % – 95 %, noncondensing
8 h
24 h
8 h
Ambient conditions, transport
• Temperature
• Air pressure
•Humidity
-10 °C to +65 °C (14 °F to 149 °F)
700 hPa – 1060 hPa
0 % – 95 %, noncondensing
Power supply: Plug-in power supply unit
• Supply voltage
• Maximum current consumption
• Insulated device, protection class II (IEC 60601-1)
12 V
Max. 1.5 A
Mains voltage 100 V – 240 V
Power supply frequency 50 Hz – 60 Hz
Power consumption max. 18 W
IEC 60601-1: Medical electrical device, Type BF
Type of protection in accordance with IEC 60529 IP 21
Duty cycle Continuous duty
Medical device in accordance with Regulation (EU) 2017/745:
• Devices without bioimpedance measurement
• Devices with bioimpedance measurement
Class I with measuring function
Class IIa
Application parts in accordance with IEC 60601-1:
Scales
seca 651/650:
Scales seca 655/654:
Scales seca mBCA 555/554, seca mBCA 552:
Standing aid seca 455:
BIA standing aid seca mBCA 550, seca mBCA 549:
Multifunctional display, glass plate
Multifunctional display, glass plate
Multifunctional display, glass plate, electrodes
Upper railing arch
Grip plates with electrodes
Interfaces:
•USB
•WiFi
•LAN
• Internal bus system/multifunctional display
USB 2.0, max. 500 mA
IEEE 802.11b/g/n/e/i
IEEE 802.3u
seca Device Bus (SDB)
Minimum weight (measurement procedure triggering on device
combinations with voice guidance)
0.5 kg