* Cannot be user set.
§ Is dependent upon codes 11 and 18.
‡ Parameter 11 plus Parameter 18 must not exceed 199.99.
Parameter
Code No.
Equivalent
Entry
Defaults/
Start-Up
Value
Unit Customer
Entered
Value
Description Enterable Range
of Values
Recommended
Range
05 Asymptote
Timer
15 secs Time delay between checks during stabilisation time 2 to 100.00 5.0 to 20.0
11‡ Range Lo 0.00 % O
2
Analogue output - zero offset -10 to 149.00 0.00 to +99.00
12 Range Hi 100.00 % O
2
* § Analogue output - zero offset + span range
13 4.00 Current output - 0 for 0-20mA, -0 for 20-0mA
- 4 for 4-20mA, -4 for 20-4mA
-4, -0, 0, 4 -4, -0, 0, 4
14 Variable % O
2
* Uncorrected oxygen input (transducer signal)
15 Variable % O
2
* Pressure input
17 Variable °C * Cell temperature
18‡ Range Span 100.00 % O
2
Span range of analogue output 1, 2, 2.5, 4, 5, 10,
20, 25, 40, 50 and
100%
1, 2, 2.5, 4, 5, 10, 20, 25, 40,
50 and 100%
19 Variable % O
2
* Calibrated oxygen reading (not compensated)
21
22
23
24
0.00
0.00
0.00
0.67
% O
2
Relay 1 of alarm board B
#
Relay 2 of alarm board #
C (if fitted)
Relay 3 of alarm board #
#
Relay 4 of alarm board D
0 to 163.80
0 to 163.80
0 to 163.80
0 to 163.80
0.40, 0.41, 0.42, 0.43,
0.45, 0.46,
0.63, 0.64, 0.65, 0.66, 0.67,
0.68,
0.92, 0.93 (and 0.62 on max,
separated versions)
0.67 is allocated
as standard to Relay 4.
25
26
27
28
29
30
Variable
secs
mins
secs
mins
secs
mins
*
*
*
*
*
*
Zero to span gas DV lag B
Span gas stabilisation time # Actual times
Span to zero gas DV lag C measured during
Zero gas stabilisation time # last auto-calibration
Zero to sample gas DV lag # sequence
Sample gas stabilisation time D
31
32
33
34
0.00
0.00
0.00
0.67
*
*
*
*
Allocatio.n of alarms to Relay 1 - dependent on code 21
Allocation of alarms to Relay 2 - dependent on code 22
Allocation of alarms to Relay 3 - dependent on code 23
Allocation of alarms to Relay 4 - dependent on code 24